Section 2: Collection of Nitrate
from Natural Waters for
Isotopic Analysis
Method I
Application: For waters with nitrate concentrations >20 micromoles
and when total DOC load is <30 mg.
Small columns: This method is suited for waters where nitrate
100 mmoles of nitrate can be collected with
<3L of water, or where DOC concentrations are such that the DOC load
of the sample does not exceed 30 mg of C. This method has been successfully
used for groundwater samples. Samples are collected on anion exchange columns
with 2 mL of resin.
Equipment: For faster, easier ordering, please see the
Resource
Guide that follows this instructional paper. The first three
items are available from BIO-RAD Laboratories.
-
Anion exchange resin AG1-X8 200-400 mesh cat.# 140-1451
-
Anion exchange columns (2 mL), cat.# 731-1550 $52/ pkg . of 50
-
Pre-filled columns are also available. cat.# 731-621 $130/pkg. of 50
-
If the sample volume is <2 liters, use Econo-Column Funnel 250 mL, cat.#731-003
$52/pkg.of 5.
-
If the sample volume is >2 liters, you will want to use a collapsable 'cube
container to hold your water sample. You will also need rubber stoppers
to fit on the top of the column and flexible tubing to connect the 'cube
container to the anion column. Drill a hole in the stopper, insert a short
piece of inflexible tubing through the stopper and use the flexible tubing
to connect the sample to the anion column. A disposable 1 mL pipet tip
with the end snipped off works nicely. Use a clamp to secure the
stopper to the column.
-
Small stopcocks that fit on the bottom of the column. Cole Parmer, cat.#
E-30600-01.
COLUMN PREPARATION
-
Rinse the resin. Put some resin in a beaker and add ~10 excess volume of
DI water. Let the resin sink to the bottom of the beaker and discard the
water. Repeat 3 times to rinse the resin. Example for 50 mL of resin, rinse
with 500 mL of water.
-
Put the columns in a rack and pack columns with 2 mL of resin.We find it
convenient to fill a lot of columns at once and then store them in the
refrigerator until we need to use them. Mark off the 2 mL level with a
pen. Pour the resin slurry into the column or use a 5 mL pipet.The
column is ready to use when the resin fills up to the 2 mL mark, after
the water has drained out of the resin.
-
If you are going to store the columns, cap the bottom of the column and
add DI water to the 6 mL mark.This keeps the resin from drying out. Cap
the top and store columns upright.
The sample should flow through the anion column no faster than 2
L per hour. Use the small stopcocks to adjust the flow rate. With
time it, may be necessary to open the stopcock in order to keep the flow
rate up.
If your sample volume is <2 L use the column funnels. Attach
the funnels to the column and slowly pour the sample through. The
funnels hold about 250 mL, so you will have to pour more sample in as the
water drips through the column.To verify that nitrate did not bleed through
the column, collect he eluant and measure for nitrate.
Method II
Large columns: This method was designed for waters where nitrate
concentrations range from 1-10 mM, necessitating
the collection of large volumes of water to collect 100 mM
of nitrate. It was originally designed for alpine waters (snow, snowmelt).
Even though the DOC concentration of these waters is low, the load of DOC
can be appreciable if 20 or more liters are needed. To accommodate
the larger volume of water and the DOC load, we use larger columns with
more resin (5 mL), a courser mesh size of resin so that the sample will
flow through faster, and we put a 5 mL cation column in front of the anion
column to minimize DOC accumulation on the anion column. To verify
that nitrate did not bleed through the column, collect the eluant and measure
for nitrate. See Nitrate Collection Poster
for a graphic example of Method II.
Equipment - Column equipment: For faster, easier ordering,
please see the Resource
Guide that follows this instructional paper. Items 1-4 available
from BIO-RAD Laboratories.
-
20 mL columns - Econo-Pac Column cat.# 732-1010, $85/pkg. of 50.
-
Fill with 5 mL of AG2X-8 anion resin (100-200 mesh), cat.# 140-2441, $210/500g
per bottle.
-
Econo-Column Funnel, 250 mL, cat.# 731-003, $52/pkg. of 5. May be
useful for loading 1 or 2 liter samples.
-
Biorad cation exchange resin- AG 50W-X8, (100-200 mesh), cat.# 142-1441,
$210/500g per bottle.
-
Stopcocks to regulate the flow coming out of the cation column. Nalgene
stopcocks, cat.# 6460-002, $111/pkg. of 6.
-
Small stopcocks for the end of the anion column. Source: Cole Parmer, cat.#
E-30600-01, $16.25/pkg. of 10. Cole Parmer, ph.1-800 323 4340.
-
Number 6 rubber stoppers with a hole.These fit on top of the columns.
Kjheldahl stoppers, Fisher cat.# 14-14913.
-
Joint clamps, stainless steel, size 35/20. Used to clamp stopper to column.
Makes a tight seal.Crimp in one side of the clamp so that it grips the
stopper. Fisher cat.# 05-8856, $151/pkg of 12.
-
Straight connectors, used to attach the columns. Daigger cat.# LX28000B,
$2.95/pkg of 10. Diager ph.1-800-621-7193. (Note: 1 mL pipet tips
with the end snipped off are also useful)
-
Nalgene tubing, 3/8 I.D., to fit onto the valve of the Cube Container.
Neoprene tubing, with 1/4 ID, used to connect the cation column to the
anion column, using a straight connector.
-
Collapsible Cube containers to put the water in when loading the columns.
-
HCL 1.2 N (10% when taking straight from a 12 N bottle) for rinsing the
plasticware used to collect and store your water samples.
-
If you have the good fortune to do this in the lab: Plastic garbage cans
(20 gal) with lids to melt water in, for processing samples.
The CAT-AG2X uses a cation column in front of the anion column. Use tubing
and straight connectors to connect the cation column to the anion column.
If the sample volume is modest, you can attach the Econo funnel to the
cation column, and pour the sample through. However, with larger sample
volumes, we put the sample in the collapsable container and let the sample
gravity drip down to the columns. If you are doing this in field,
simply place the container on a ledge above the columns and let the sample
pass slowly through the cation and anion columns.
DOC equipment
-
Silver and GF/F filters for DOC filtration. Filters must be combusted prior
to filtration.(see step 2, in loading the columns). Glass fiber filters
serve as a prefilter.Order from PORETICS,111 A Lindbergh Ave. Livermore,
CA, 94550. Ph: 1-800-922-6090. Silver Filters, cat.# 53014(0.2 micron,47mm
diameter),$244.80/pkg. of 100 (government pricing). Also available in pkgs.of
25.GF/F type filters, glass fiber, cat.# 88231, $21./pkg.of 100.Combust
silver filters at 200 C for half an hour. Do not over bake. Note
that there is a wrong and right side, dull side up! Combust glass filters
2 hours at 400 C.Combust filters in a glass container, a beaker covered
with tin foil or a civered glass petri dish . Store in said containers.
-
Filtration unit that will accept 47 mm diameter filters. Unit should be
stainless steel or Teflon, to avoid contamination that can be introduced
with a glass unit.
-
Glass bottles for the DOC sample. Boston round bottles that hold 100 mL
are good.Obtain bottles with Teflon cap liners. Remove the cap and place
a piece of tin foil on the top of the bottle.Combust the bottles with the
foil tops in a furnace, 400 C for 2 hours. Do not combust the caps
as they will melt.
-
Combust about twenty, one and a half inch squares, of aluminum foil in
a glass container as in step 1.Combust with bottles.
-
Filter forceps.
To determine nitrate concentrations:
Use a Hach Kit to make field measurements or have your samples analyzed
by a lab. If you decide to have your samples analyzed by a lab, you must
take a small sample.
-
HDLPE bottles, 100 mL.One bottle for each sample.
Filtering equipment
If your nitrate concentrations are high enough so that you need
only a few liters of water, then you can filter your sample with a disk
filter. Some people like to use the large,147 mm filters (0.45
mm).
Use a nylon filter or a polycarbonate filter. Do not use a filter made
of cellulose nitrate. If the filter clogs, you can use a GF/F type
glass fiber filter (0.7mm) as a prefilter. Put
the glass filter on top of the 0.45mm filter.
The filter must not have a binder in it. Poretics makes a GF/F type filter
for a more affordable price than Whatman. If you have a lot of material
in your sample, then use a glass fiber filter with a larger pore size,
(1 or 2 mm) but be sure it does not contain
a binder. If you must pass large volumes of water (>20 L) through a column,
because your NO3 concentration is low, use the Gelman groundwater
capsule. Filtration is rapid, thereby saving you a lot of time.
-
Gelman "groundwater sampling capsules" (0.45m
m filter) polystyrene, cat.# 12176, $150/for a box of 10. Order from
Fisher Scientific.
-
Peristaltic pump, used with the capsules.
-
Silicone or neoprene tubing, 3/8" ID, with walls that are 1/16 or 1/32
for the peristaltic pump or some type of tubing that won' t adsorb or disturb
nitrate. Necessary only if you are using the Gelman capsules.
-
Ring stands and clamps. Support rack for columns.
-
Vacuum Pressure bulb, Fisher cat.#14-080.
Information needed before the columns can be loaded.
Determine your nitrate concentrations!
-
Rinse each plastic bottle, the collapsible containers, and garbage cans,
in the HCL solution.Rinse three times with DI water.If you are collecting
a small sample for nitrate determinations, then the small sample bottles
must be washed and rinsed too.With the smaller bottles, we advise that
you fill them with DI and let them soak overnight.
-
Currently, we are requesting a minimum of a 100 mmoles
of nitrate on each column.To determine the volume of sample required to
flow through the column, you must know the nitrate concentration in each
of you samples.To do this, filter the sample through a 0.45 micron nylon
or polycarbonate filter to remove particulates.
Rinse your sample bottle with a few mL of your sample first,
to condition the surface of the bottle, discard, and then fill with your
sample. Fill you small sample bottle (100 mL) with your sample.Use this
sample to determine your nitrate concentration and calculate the volume
of water you will need to load 100 mmoles of
nitrate. If nitrate concentrations are low, 2 mM
or less, it may be too difficult to obtain 100 mmoles
nitrate. In this case, collect as much as you can.The sample will probably
be analyzed for N-15 only. Record the nitrate concentration, liters of
sample passed through the column, and mass of nitrate loaded onto each
column in the spreadsheet. Do this for each sample.
-
Determine your Cl, SO4 and PO4 concentrations and
calculate the load in micromoles (concentration times vol) that will be
loaded onto the anion column.These can interfere with our analysis. We
have to remove them, before the sample can be analyzed. Estimate
these concentrations, if you don't know them. Enter this data in the appropriate
place on the spreadsheet. Note, that the masses of these constituents,
will be calculated on the spreadsheet once you enter the appropriate information.
One mL of resin will absorb 1.2 meq of anions, so 5 mL is 6 meq or 6000
mmoles.
If only nitrate were absorbed, the column could hold 6000 mmoles
of nitrate. Remember, the column will adsorb all anions. If your samples
contains large quantities of other anions, Cl in particular and DOC, you
will have to use more than one column to obtain the required 100 mmoles
of nitrate.
-
Determine DOC concentrations. The DOC concentrations of the sample, along
with the mass in mg loaded onto the column, should be known.This is because
dissolved organic carbon (DOC) can adsorbing onto the sites that the nitrate
would adsorb to, potentially clogging the anion column. This would
lower the ability of the column to retain nitrate, and possibly interfere
with the d18O.
Filter about 100 mL of your sample through the silver filter,
directly into the combusted glass bottle. (Rinse the bottle out with a
few mL of your sample, to condition the bottle. Discard the sample and
fill). If your sample seems to clog the silver filter, put a glass fiber
filter on top, as a prefilter. Remember, the silver filter goes dull
side up. Save the silver filters, and combust again for reuse. Replace
the foil and screw the lid on. Refrigerate and send to a lab for DOC analysis.
Do
not freeze. The silver acts as a bactericide. If your samples
can be analyzed within 24 hours, they can be filtered just with the glass
filter. Enter the information about DOC on the spreadsheet.
LOADING THE SAMPLE
-
Filter the sample. If you have a lot of samples and/or large volumes, store
your water in a cool place to discourage bacterial growth.
-
Put the sample in a cube container.
-
Note, the sample contains gases that can cause bubbles to form in the resin.
Bubbles slow the flow through the resin and reduce your efficiency. Gases
can be removed by allowing the sample to come to room temperature. We have
found that stirring the sample for half an hour, under a vacuum, helps
to remove gas. Insert a piece of tubing through the stopper hole. Put a
piece of saran wrap or parafilm over the stopper so that the sample will
not contact the rubber directly. Attach tubing from the container spigot,
to the tubing, in the stopper. Push the stopper down firmly so that it
makes a tight seal, clamp down.
-
Insert a stopper with a hole, into the cation and anion columns.To connect
the cation column to the anion column, use a straight connector and a piece
of 1/4 ID tubing. Connect a short piece (7 cm) of tubing to the bottom
of the cation column and insert a straight connector to the other side
of the tubing. Insert the straight connecter into the stopper hole of the
anion column.
-
Attach a piece of 3/8 inch, ID tubing, to the spigot of the cube container.
-
Connect the cation column to the tubing using your connectors or Nalgene
stopcocks. Remember the longer the head distance, the more pressure there
will be on the column, and the water will flow through more quickly.
-
Attach the small stopcock on the bottom of the anion column to regulate
the outflow.
-
Stabilize the column (use a rack or ring stand with clamps) to prevent
it from swaying. If you are using large volumes of water, you may want
to put a large stopcock between the columns so that you can stop the flow
from the cation column, when you need to put on a fresh anion column.
-
Collect the eluent and measure the nitrate concentration. You want to verify
that there was no bleed-out of nitrate.
-
Fill each column with about 5 mL of DI water or with the sample, to prime
the flow. Otherwise, when you open the stopcock of the cube container,
it will be difficult for the sample to flow through (back pressure).
-
Allow the sample to flow through the cation and anion columns. As the sample
flows through the columns, bubbles may develop in the resin. Insert a small
glass rod or a straightened paper clip and tease out the bubbles. Bubbles
cause the water to move around the them, causing sample flow to slow down,
thus decreasing the efficiency of the resins ability to absorb nitrate.
Do not stir the resin or shake the column!
-
Gravity drip the sample through. We think a rate of 2 L per hour is good.
Make a note of how fast your samples drip.Whether the sample can drip faster,
has yet to be determined. If the flow rate is too fast, use the stopcock
to slow the flow. As the sites fill, the flow rate will slow. Adjust as
needed.
-
As the sample is loaded on the column, the resin will begin to discolor.When
all the resin is discolored or when the column ceases to drip, change to
a new anion column. Discoloration usually indicates a DOC problem. Remember,
you may need to use more than one column per sample to obtain your 100
µmoles of nitrate.
-
After samples have passed through the column(s), blow or vacuum excess
water from column. Fill the column to the top with DI water, cap top and
bottom and refrigerate. Seal the top and bottom with parafilm, so they
don't open upon shipping. A one-way bulb attached to an appropriately sized
stopper works well to blow the columns dry.
STORAGE AND SHIPPING
-
Store columns refrigerated.
-
When shipping, place columns in double "Zip-Lock," plastic bags and ship
in coolers.
-
That's it! You are finished. Send the columns and your spreadsheet
(Quattro Pro or Excel) to us (see instructions for spreadsheet Sampledata.htm).
Include a hard copy of the file, as well as a disk, when you send us the
information.
- IMPORTANT NOTICE -
The use of name brands through out this
document is for identification purposes only and does not constitute any
endorsements by the U.S. Geological Survey
RESOURCE GUIDE
-
BIO-RAD Laboratories, 200 Alfred Nobel Drive, Hercules,
CA 94547, PH:1-(800)424-6723. BIO-RAD
-
Cole-Parmer Instrument Co., 625 East Bunker Court,
Vernon Hills, IL 60061-1844, PH: 1-800-323-4340. COLE-PARMER
-
Daigger, 675 Heathrow Drive, Lincolnshire, IL 60069-4206,PH:1-800-621-7193.
DAIGGER
-
Fisher Scientific, U.S. Headquarters, 585 Alpha Drive,
Pittsburg, PA 15238, PH: 1-800-766-7000. FISHER
-
Gelman Sciences, 600 Wagner Road, An Arbor, MI 48103-9019,PH:
1-800-521-1520. GELMAN
SCIENCES
-
HACH Company, World Headquarters, P.O. Box 389, Loveland,
CO 80539-0389, PH: 1-800-227-4224. HACH
-
Poretics Corporation, Order Service, 111 Lindbergh
Avenue, Livermore, CA 94550-9261, PH: 1-800-922-6090. PORETICS
-
Whatman LabSales, P.O. Box 1359, Hillsboro, OR 97123-9881,
PH: 1-800-942-8626. WHATMAN
Return to Index
The URL of this page is: http://wwwrcamnl.wr.usgs.gov/isoig/guidelines/nitrate/collection.htm
Page created by Carla Chamberlain, Daniel Steinitz
and Cecily Chang.
Please contact Carol Kendall (ckendall@usgs.gov)
for questions and comments regarding this page.
This page was last changed on July 19, 2001.
To
the Isotope Interest Group Home Page
To the USGS Home
Page
To the Water Resources
Information Home Page
|